Features:
- The innovative optical layout and state of art mono- chromator with high grade blazed holographic grating ensure accuracy. Its integrated design assures long term stability and durability
- The precisely aligned detector and quality deuterium and halogen lamps enhance the precision across the UV / Vis spectrum starting from 190 nm and into the near infrared 1100nm
- The two detectors are used to measure sample and reference respectively and simultaneously for optimizing measurement accuracy
- Lamp Selection enables conserving the life of D2 & Tungsten lamps. Real time clock for date & time stamping of results
- Big LCD display. (320 x 240 Dots Matrix)
- Basic Mode, Quantitative, Kinetics, Multi – Wavelength, Wavelength Scan, DNA / Protein Test &Utility
Function Of Software:
- Basic Mode
- To measure the Absorbance and Transmittance
Quantitative:
- Coefficient Method
- Standard Method curve.
- Up to 10 standard samples may be used to establish a curve. Four methods to fitting a curve through the calibration points: Linear fit through zero, square fit & cubic fit.
Wavelength Scan:
- . The wavelength scan intervals are 0.1, 0.2, 0.5, 1, 2 & 5 nm
- 2. Fast, Medium & Slow Scan speeds are available. They very from100 to 1800 nm/min
- 3. Wavelength are scanned from high to low so that the instrument waits at high Wavelength and minimizes the degradation of UV sensitive samples
Multi-Wavelength:
- Up to 32 wavelengths can be selected and multiple samples can be measured
Kinetics:
- This mode may be used for time course scanning or reaction rate calculations. Abs. vs. time graphs is displayed on the screen in real time
- Wait time and measurement time up to 12 hours may be entered with time intervals of 0.5,1,2,5,10,30 seconds & 1min. Post- run manipulation includes re-scaling, curve tracking and selection of the part of the curve required for the rate calculation. Rate is calculated using a linear regression algorithm before multiplying by the entered factor
DNA/Protein Test:
- Concentration and DNA purity are quickly and easily calculated: Absorbance ratio s 260nm/280nm with optional subtracted absorbance at 320nm.
- DNA concentration= 62.9x A260-36.0x A280
- Protein concentration=1552xA260-757.3xA280
Utility:
- Has following functions
- WL Reset, Printer, Lamp, Clock, Dark Current, Accu. Validity, WL Validity Connect to PC, Beeper On/Off & System Base Line
Labman Portable Microprocessor UV Visible Spectrophotometer
Wavelength Range 190 nm Spectral Bandwidth 1 nm Mobility Portable Type Portable Display 320 x 240 Graphic LCD Optical System Double Beam, Grating 1200 lines / mm Brand Labman Model Name/Number LMSP-UV1900 Photometric Accuracy +0.3%T Photometric Range -0.3-3A, 0-200%T, 0-9999C Photometric Repeatability + 0.2%T Data Output Port USB Cell Holder 10mm single cuvette Holder Dimensions 625 x 430 x 206 mm Printer Port Parallel Port Wavelength Repeatability + 0.2 nm Material ABS Plastic Wavelength Accuracy + 0.3nm Stray Light <0.05%T @ 220nm, 360nm Stability + 0.001A/h @ 500nm Lamps Deuterium Lamp & Tungsten Halogen Lamp Weight 30 kg Detector Silicon Photodiode Bandwidth 1nm Scanning Speed Fast, Medium & Low Baseline Flatness & Noise +0.001A Power Supply AC 220V/50Hz